Glucagon Kit HTRF®

The Glucagon kit is intended to provide significant time-saving over ELISA, in the quantification of glucagon in cell culture media.
See more
  • Low sample consumption Low sample consumption
  • Faster and more convenient than ELISA Faster and more convenient than ELISA
  • Scalable from 96 to 1536 format Scalable from 96 to 1536 format
The Glucagon kit is intended to provide significant time-saving over ELISA, in the quantification of glucagon in cell culture media.
-

Overview

The Glucagon assay offers superior sensitivity and ease-of-use for a reliable quantification of glucagon in cell culture media. Built on a truly homogeneous protocol, the assay does not require any washing or complex extraction steps, meaning significant time savings over ELISA methods. Simply add the detection reagents, incubate, and read!

Benefits

  • HIGH SPECIFICITY

Assay principle

Glucagon is measured using a sandwich immunoassay with two monoclonal antibodies, one labelled with Lumi4-Tb Cryptate (donor) and the other with d2 (acceptor). The intensity of the FRET signal obtained is proportional to the concentration of glucagon in the sample.
Glucagon kit principle

Assay protocol

10 µL of cell supernatants are supplemented with 5 µL of acceptor-labelled antibodies and 5 µL of donor-labelled antibodies (or 10 µL of pre-mixed reagents for a single dispensing step). The mix is then incubated from 4h to overnight at RT and results are read on an HTRF compatible reader.
Glucagon kit protocol

Specificity

PeptideSpecificity (% of recognition)
Glucagon1
Oxyntomodulin<0.07%
Glicentin<0.07%
Glucagon fragment 1-18<1.81%
Glucagon fragment 19-29<0.03%
GLP-1 (7-36) amide<0.06%
GLP-1 (7-37)<0.11%
GLP-2<0.3%
GRPP (Glicentin-Related Pancreatic Peptide)<0.01%

Sensitivity

Typical calibration curves obtained by the dilution of glucagon standard in cell-culture media conditions for 20 µL final, using a white 384-well small volume plate, are presented below. Samples were incubated overnight at RT and diluted in diluent #5. The assay ranges from 15,6 to 2000 pg/mL and the detection limit measured was 6 pg/mL (PHERAstar FS reader - flash lamp excitation).
Glucagon standard curve with serial dilution in diluent #5

Linearity

Dilution linearity in cell culture media was assessed. Recovery results of between 88% and 100% were obtained, supporting the claims regarding linearity. Mouse a-TC1-6 cells were seeded in 96-well culture treated plates (100k cells/well) in complete culture medium containing 25mM glucose and incubated for 24 h at 37°C - 5% CO2. The media was then removed and the wells washed twice with KRB buffer before being incubated 1 h with KRB buffer containing low glucose (1 mM). Supernatants were collected and serially diluted in diluent #5. Results are expressed as percent observed from expected.
Glucagon kit linearity curve with serial dilutions of mouse cell supernatant sample in diluent #5
Serial dilutions of a mouse a-cell supernatant in diluent #5: Mouse a-cell supernatant [Glucagon] measured (pg/mL) [Glucagon] expected (pg/mL) % of expected concentration Undiluted 1460.0 - - 1:2 733.0 730.0 100.4 % 1:4 351.0 365.0 96.2 % 1:8 167.0 182.5 91.5 % 1:16 80.3 91.3 88.0 %

Intra- and inter-assay variability

Intra-assay and inter-assay variability in cell culture media conditions were assessed. Typical CVs of less than 5% were obtained, supporting the features required for a robust and reproducible assay.

Intra-Assay - n = 24

CV

Standard 2 (= 31.25 pg/mL)

2.8%

Standard 6 (= 500 pg/mL)

2.4%

Standard 8 (= 2,000 pg/mL)

3.0%

Inter-Assay - n = 6

CV

Standard 2 (= 31.25 pg/mL)

8.1%

Standard 6 (= 500 pg/mL)

4.4%

Standard 8 (= 2,000 pg/mL)

5.4%

Validation on mouse isolated pancreatic islets

15 pancreatic islets isolated from a C57BL/6J mouse were exposed to increasing concentrations of Glucose or Adrenaline. Following exposure, supernatants were collected and their Glucagon levels quantified. In the presence of glucose, Glucagon levels were readily detected using as few as 15 islets per well. All results were similar to those previously published by other investigators. Courtesy of Physiopathology of pancreatic BETA cell research team, Institute de Génomique Fonctionnelle, Montpellier, France.
Quantification of glucagon in mouse pancreatic islets for different glucose concentrations

Product Insert Glucagon Kit / 62CGLPEG-62CGLPEH

62CGLPEG-62CGLPEH - Product Insert

HTRF Product Catalog

All your HTRF assays in one document! - Catalog

A guide to Homogeneous Time Resolved Fluorescence

General principles of HTRF - Guides

A guide to help you in your diabetes projects

A comprehensive listing of Type II diabetes pathways - Guides

How HTRF compares to Western Blot and ELISA

Get the brochure about technology comparison. - Brochures

Plate Reader Requirement

Choosing the right microplate reader ensures you’ll get an optimal readout. Discover our high performance reader, or verify if your lab equipment is going to be compatible with this detection technology.

Let's find your reader